Chemical Sciences: A Manual for CSIR-UGC National Eligibility Test for Lectureship and JRF/Accelerator mass spectrometry

From Wikibooks, open books for an open world
Jump to navigation Jump to search


Accelerator mass spectrometry (AMS) differs from other forms of mass spectrometry in that it accelerates ions to extraordinarily high kinetic energies before mass analysis. The special strength of AMS among the mass spectrometric methods is its power to separate a rare isotope from an abundant neighboring mass ("abundance sensitivity", e.g. 14C from 12C).[1] The method suppresses molecular isobars completely and in many cases can separate atomic isobars (e.g. 14N from 14C) also. This makes possible the detection of naturally occurring, long-lived radio-isotopes such as 10Be, 36Cl, 26Al and 14C. Their typical isotopic abundance ranges from 10-12 to 10-18. AMS can outperform the competing technique of decay counting for all isotopes where the half life is long enough.[2]

The method

Generally, negative ions are created (atoms are ionized) in an ion source. In fortunate cases this already allows the suppression of an unwanted isobar, which does not form negative ions (as 14N in the case of 14C measurements). The pre-accelerated ions are usually separated by a first mass spectrometer of sector-field type and enter an electrostatic "tandem accelerator". This is a large nuclear particle accelerator based on the principle of a Tandem van de Graaff Accelerator operating at 0.2 to many million volts with two stages operating in tandem to accelerate the particles. At the connecting point between the two stages, the ions change charge from negative to positive by passing through a thin layer of matter ("stripping", either gas or a thin carbon foil). Molecules will break apart in this stripping stage.[3][4] The complete suppression of molecular isobars (e.g. 13CH- in the case of 14C measurements) is one reason for the exceptional abundance sensitivity of AMS. Additionally, the impact strips off several of the ion's electrons, converting it into a positively charged ion. In the second half of the accelerator the now positively charged ion is accelerated away from the highly positive center of the electrostatic accelerator which previously attracted the negative ion. When the ions leave the accelerator they are positively charged and are moving at several percent of the speed of light. In a second stage of mass spectrometer, the fragments from the molecules are separated from the ions of interest. This spectrometer may exist of magnetic or electric Sector instrument|sectors, and so called velocity selectors, which utilizes both electric fields and magnetic fields. After this state, no background is left, unless a stable (atomic) isobar forming negative ions exists (e.g. 36S if measuring 36Cl), which is not suppressed at all by the setup described so far. Thanks to the high energy of the ions, these can be separated by methods borrowed from nuclear physics, like degrader foils and gas-filled magnets. Individual ions are finally detected by single-ion counting (with silicon surface-barrier detectors, ionization chambers, and/or time-of-flight telescopes). Thanks to the high energy of the ions, these detectors can provide additional identification of background isobars by nuclear-charge determination.

Generalizations

The above is an example. There are other ways in which AMS is achieved; however, they all work based on improving mass selectivity and specificity by creating high kinetic energies before molecule destruction by stripping, followed by single-ion counting.

History

Luis Walter Alvarez|L.W. Alvarez and Robert Cornog of the United States first used an accelerator as a mass spectrometer in 1939 when they employed a cyclotron to demonstrate that 3He was stable; from this observation, they immediately (and correctly) concluded that the other mass-3 isotope tritium was radioactive. In 1977, inspired by this early work, Richard A. Muller at the Lawrence Berkeley Laboratory recognized that modern accelerators could accelerate radioactive particles to an energy where the background interferences could be separated using particle identification techniques. He published the seminal paper in Science (vol 196, pages 489-494, 1977) showing how accelerators (cyclotrons and linear) could be used for detection of tritium, radiocarbon (14C), and several other isotopes of scientific interest including 10Be; he also reported the first successful radioisotope date experimentally obtained using tritium (3H). His paper was the direct inspiration for other groups using cyclotrons (G. Raisbeck and F. Yiou, in France) and tandem linear accelerators (D. Nelson, R. Korteling, W. Stott at McMaster). K. Purser and colleagues also published the successful detection of radiocarbon using their tandem at Rochester. Soon afterwards the Berkeley and French teams reported the successful detection of 10Be, an isotope widely used in geology. Soon the accelerator technique, because it was about a factor of 1000 more sensitive, virtually supplanted the older “decay counting” methods for these and other radioisotopes.

Applications

The applications are many. AMS is most often employed to determine the concentration of carbon-14|14C, e.g. by. Archaeologists for radiocarbon dating. An accelerator mass spectrometer is required, over other forms of mass spectrometry, because of their insufficient abundance sensitivity, and to resolve stable nitrogen-14 from radiocarbon. Due to the long half-life of 14C, decay counting requires significantly larger samples. 10Be, 26Al, and 36Cl are used for surface exposure dating in geology. 3H, 14C, 36Cl, and 129I are used as hydrological tracer.

Accelerator mass spectrometry is widely used in biomedical research.[5][6][7]

References

  1. "abundance sensitivity (in mass spectrometry)" (PDF), IUPAC Compendium of Chemical Terminology, International Union of Pure and Applied Chemistry {{citation}}: Unknown parameter |chapterurl= ignored (|chapter-url= suggested) (help)
  2. Budzikiewicz H, Grigsby RD (2006). "Mass spectrometry and isotopes: a century of research and discussion". Mass spectrometry reviews. 25 (1): 146–57. doi:10.1002/mas.20061. PMID 16134128.
  3. Litherland, A. E. (1980). "Ultrasensitive Mass Spectrometry with Accelerators". Annual Review of Nuclear and Particle Science. 30: 437–473. doi:10.1146/annurev.ns.30.120180.002253.
  4. R. d. L. John (1998). "Mass spectrometry and geochronology". Mass Spectrometry Reviews. 17 (2): 97–125. doi:10.1002/(SICI)1098-2787(1998)17:2<97::AID-MAS2>3.0.CO;2-J.
  5. Brown K, Dingley KH, Turteltaub KW (2005). "Accelerator mass spectrometry for biomedical research". Meth. Enzymol. 402: 423–43. doi:10.1016/S0076-6879(05)02014-8. PMID 16401518.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  6. Vogel JS (2005). "Accelerator mass spectrometry for quantitative in vivo tracing". BioTechniques. Suppl: 25–9. doi:10.2144/05386SU04. PMID 16528913.
  7. Palmblad M, Buchholz BA, Hillegonds DJ, Vogel JS (2005). "Neuroscience and accelerator mass spectrometry". Journal of mass spectrometry : JMS. 40 (2): 154–9. doi:10.1002/jms.734. PMID 15706618.{{cite journal}}: CS1 maint: multiple names: authors list (link)

Bibliography

  • edited by J.A.J. Gowlett and R.E.M. Hedges. (1986). Archaeological results from accelerator dating: research contributions drawing on radiocarbon dates produced by the Oxford Radiocarbon Accelerator based on papers presented at the SERC sponsored conference Results and prospects of accelerator dating held in Oxford on October 1985. Oxford: Oxford University Committee for Archaeology. ISBN 0-947816-11-9. {{cite book}}: |author= has generic name (help)
  • Gove, H. E. (1999). From Hiroshima to the iceman: the development and applications of accelerator mass spectrometry. London: Institute of Physics. ISBN 0-7503-0557-6.
  • Tuniz, C. (1998). Accelerator mass spectrometry: ultrasensitive analysis for global science. Boca Raton: CRC Press. ISBN 0-8493-4538-3.